Agilent B1500A/B1505A Programming Guide, Edition 11 2-63
Remote Mode Functions
SPGU Module
Table 2-5 ALWG Sequence Data (binary format, big endian)
In the example of Figure 2-34, the total byte length of the pattern data is as follows.
• SPGU 801 number of patterns: x=2 (Pattern1 and Pattern2)
Number of vectors for Pattern1 and Pattern2: N
1
=9, N
2
=9
Total byte length is 20+(6+8×9)+(6+8×9)=176 bytes.
• SPGU 901 number of patterns: x=2 (Pattern1 and Pattern2)
Number of vectors for Pattern1 and Pattern2: N
1
=9, N
2
=7
Total byte length is 20+(6+8×9)+(6+8×7)=160 bytes.
And, the total byte length of the sequence data is as follows.
Number of pattern cycles: x=2 (Pattern1 and Pattern2)
Total byte length is 20+6×2=32 bytes.
Data Data length Byte length
Header Module type (ex: 0) 1 byte 20 bytes
Data format revision (ex: 0) 1 byte
Number of pattern cycles
a
(ex: x)
a. Number of pattern cycles must be 1 to 1000000000 (512).
2 bytes
Others (0 for all bit. Do not change.) 16 bytes
Sequence
data
Pattern
cycle data
Pattern index (ex: 1 for Pattern1) 2 bytes 6× i bytes,
i=1 to x
Repeat count
b
(ex: 5)
b. Repeat count must be 1 to 100000000000000000000 (1048576).
4 bytes
:
Pattern cycle data can be repeated until the x-th pattern cycle.
(x: number of pattern cycles, integer)