Agilent Technologies B1500A Manual de servicio Pagina 283

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 598
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 282
Agilent B1500A/B1505A Programming Guide, Edition 11 3-105
Programming Examples
SPGU Pulse Output and Voltage Measurement
SPGU Pulse Output and Voltage Measurement
To control the SPGU channel, use the following commands.
Function Command Parameters
Enables channels CN [chnum ... [,chnum] ... ]
Disables channels CL [chnum ... [,chnum] ... ]
Sets SPGU pulse output mode SIM 0
Sets output operation mode SPRM mode[,condition]
Sets pulse period for all of
SPGU channels
SPPER period
Sets channel output mode SPM chnum,mode
Sets pulse timing parameters SPT chnum,src,delay,width,leading
[,trailing]
Sets pulse level parameters SPV chnum,src,base[,peak]
Sets load impedance SER chnum,loadZ
Measures terminal voltage,
calculates and sets load
impedance
[CORRSER?] chnum,mode,delay,interval,
count
Sets pulse switch [ODSW] chnum,state[,normal[,delay,
width]]
Sets SPGU trigger output [STGP] chnum,state
Applies SPGU channel setup SPUPD chnum[,chnum ... [,chnum] ... ]
Starts SPGU pulse output SRP
Stops SPGU pulse output [SPP]
Returns SPGU output status [SPST?]
Vista de pagina 282