5-16 Agilent B1500A/B1505A Programming Guide, Edition 11
Error Messages
Operation Error
696 HVSMU Current Expander control cable was disconnected.
Connection cable was removed. Turn the instrument off and connect
the cable. And then turn the instrument on again.
697 HVSMU Current Expander is not active.
HVSMU current expander does not respond. The power switch may be
off or the power code may be removed.
940 DIO control mode must be Ultra High Voltage Expander control mode
(ERMOD 16).
Set the digital IO control mode to N1268A control mode by using the
ERMOD 16 command.
941 Mismatch Ultra High Voltage Expander ID.
Ultra high voltage expander is not connected to the Digital IO
connector. Connect the N1268A.
942 Ultra High Voltage Expander control cable was disconnected.
Connection cable was removed. Turn the instrument off and connect
the cable. And then turn the instrument on again.
943 Ultra High Voltage Expander is not active.
Ultra high voltage expander does not respond. The power code may be
removed.
999 Error not supported by the “ERR?” command occurs. Execute the
“ERRX?” command to read the error code and the error message.
1000 The specified module doesn’t support power compliance.
The power compliance is not available for the specified module.
1001 Illegal pulse duty
Set the pulse period and the pulse width so that the pulse duty ratio is
within the acceptable range.
1002 Illegal pulse width
Check the pulse width and set the correct value.
1003 Illegal pulse base/peak limit
Check the pulse base and peak values, and set the correct values.