4-232 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
WTDCV
Tdelay : Step source trigger delay time (in seconds) that is the wait time after
completing a step output setup and before sending a step output setup
completion trigger.
0 (initial setting) to delay or 65.535, with 0.1 ms resolution. Numeric
expression. If this parameter is not set, Tdelay will be 0.
Mdelay : Step measurement trigger delay time (in seconds) that is the wait time
after receiving a start step measurement trigger and before starting a
step measurement.
0 (initial setting) to 65.535, with 0.1 ms resolution. Numeric
expression. If this parameter is not set, Mdelay will be 0.
Example
Statements
OUTPUT @B1500;"WTACV 5,0.1,0.1,0.1,0.1"
OUTPUT @B1500;"WTACV 5,0.2"
WTDCV
This command sets the hold time, delay time, and step delay time for the CV (DC
bias) sweep measurement (MM18). This command is also used to set the step source
trigger delay time effective for the step output setup completion trigger and the step
measurement trigger delay time effective for the start step measurement trigger. For
the trigger function, refer to “Trigger Function” on page 2-74. If you do not enter
this command, all parameters are set to 0.
Syntax WTDCV hold,delay[,Sdelay[,Tdelay[,Mdelay]]]
Parameters hold : Hold time (in seconds) that is the wait time after starting measurement
and before starting delay time for the first step.
0 to 655.35, with 10 ms resolution. Numeric expression.
delay : Delay time (in seconds) that is the wait time after starting to force a step
output and before starting a step measurement.
0 to 65.535, with 0.1 ms resolution. Numeric expression.
Sdelay : Step delay time (in seconds) that is the wait time after starting a step
measurement and before starting to force the next step output.
0 to 1, with 0.1 ms resolution. Numeric expression.
If this parameter is not set, Sdelay will be 0.
If Sdelay is shorter than the measurement time, the B1500 waits until
the measurement completes, then forces the next step output.