Agilent B1500A/B1505A Programming Guide, Edition 11 4-37
Command Reference
ADJ
0 mV (initial setting) to 250 mV, 1 mV step.
Example
Statements
OUTPUT @B1500;"ACV 7,0.01"
ADJ
This command selects the MFCMU phase compensation mode. This command
initializes the MFCMU.
Syntax ADJ chnum,mode
Parameters chnum : MFCMU channel number. Integer expression. 1 to 10 or 101 to 1001.
See Table 4-1 on page 4-16.
mode : Phase compensation mode. Integer expression. 0 or 1.
0: Auto mode. Initial setting.
1: Manual mode.
For mode=0, the B1500 sets the compensation data automatically.
For mode=1, execute the ADJ? command to perform the phase
compensation and set the compensation data.
Example
Statements
OUTPUT @B1500;"ADJ 9,1"
ADJ?
This command performs the MFCMU phase compensation, and sets the
compensation data to the B1500. This command also returns the execution results.
This command resets the MFCMU.
Before executing this command, set the phase compensation mode to manual by
using the ADJ command. During this command, open the measurement terminals at
the end of the device side. This command execution will take about 30 seconds.
The compensation data is cleared by turning the B1500 off.
Syntax ADJ? chnum[,mode]
Parameters chnum : MFCMU channel number. Integer expression. 1 to 10 or 101 to 1001.
See Table 4-1 on page 4-16.
mode : Command operation mode.
0: Use the last phase compensation data without measurement.