Agilent Technologies B1500A Manual de servicio Pagina 511

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 598
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 510
Agilent B1500A/B1505A Programming Guide, Edition 11 4-195
Command Reference
TGMO
TGMO
The TGMO command selects the edge trigger or the gate trigger for the Step
Measurement Completion trigger output set by the TGP port,2,polarity,3
command. See Figure 4-4.
This command is available for the staircase sweep, multi channel sweep, and
MFCMU DC/AC/frequency sweep measurements.
Figure 4-4 Trigger Output Example, Staircase Sweep Measurement, Negative Logic
Syntax TGMO mode
Parameters mode : Edge trigger or gate trigger. Integer expression. 1 or 2.
1: Edge trigger (initial setting).
2: Gate trigger.
Example
Statements
OUTPUT @B1500;"TGMO 2"
See Also “TGP” and “TGPC”
: Measurement
TGP t,2,p,3
XE
TGP t,2,p,2
TGP t,2,p,1
TGP t,2,p,3
TGP t,2,p,2
TGP t,2,p,1
Gate trigger
Edge trigger
Source trigger delay
Hold time
Delay time
Source trigger delay
Step delay time
Step Measurement Completion
Step Output Setup Completion
Measurement Completion
Step Measurement Completion
Step Output Setup Completion
Measurement Completion
Vista de pagina 510
1 2 ... 506 507 508 509 510 511 512 513 514 515 516 ... 597 598

Comentarios a estos manuales

Sin comentarios