Agilent B1500A/B1505A Programming Guide, Edition 11 4-201
Command Reference
TI
TI
The TI command performs the high speed spot measurement, and returns the
measurement data. The command starts a current measurement regardless of the
SMU operation mode, trigger mode (TM command), and measurement mode (MM
command).
Execution
Conditions
The CN/CNX command has been executed for the specified channel.
Syntax TI chnum[,range]
Parameters chnum : SMU measurement channel number. Integer expression. 1 to 10 or 101
to 1001. See Table 4-1 on page 4-16.
range : Measurement range or ranging type. Integer expression. See Table 4-3
on page 4-19.
If you do not specify the range parameter for voltage output channels, the channel
uses the minimum range that covers the compliance value.
Example
Statements
OUTPUT @B1500;"TI 1"
ENTER @B1500 USING "#,3X,13D,X";Idata
TIV
The TIV command performs the high speed spot measurement, and returns the
measurement data. The command starts a current and voltage measurement
regardless of the SMU operation mode, trigger mode (TM command), and
measurement mode (MM command).
The HCSMU and HVSMU perform the current measurement and the voltage
measurement simultaneously.
The HRSMU, MPSMU, and HPSMU perform the compliance side measurement
and the force side measurement in this order.
Execution
Conditions
The CN/CNX command has been executed for the specified channel.
Syntax TIV chnum[,irange,vrange]
Parameters chnum : SMU measurement channel number. Integer expression. 1 to 10 or 101
to 1001. See Table 4-1 on page 4-16.