2-52 Agilent B1500A/B1505A Programming Guide, Edition 11
Remote Mode Functions
Dual HCSMU
Dual HCSMU
This function is available when two HCSMU modules are installed in the B1505A
and connected to the 16493S-020 Dual HCSMU Kelvin combination adapter or the
16493S-021 Dual HCSMU combination adapter. Two HCSMU modules can
perform the dual HCSMU operation which supports ± 40 A (pulse), ± 2 A (DC).
The dual HCSMU operation is enabled by the following command.
PCH master,slave
where,
master : Channel number of HCSMU used as the dual HCSMU master channel
slave : Channel number of HCSMU used as the dual HCSMU slave channel
After the PCH command is executed, the HCSMU modules of channel numbers
master and slave operate as the dual HCSMU (DHCSMU) of channel number
master, and do not perform the single HCSMU operation.
The specified channels must have been disabled when the PCH command is
executed. Do not execute the CN/CNX command for the specified channels before
the PCH command.
NOTE Available output and measurement ranges
The following output and measurement ranges are available for the dual HCSMU.
Current ranges: 10 μA, 100 μA, 1 mA, 10 mA, 100 mA, 1 A, 2 A, and 40 A
Voltage ranges: 0.2 V, 2 V, 20 V, and 40 V
The dual HCSMU does not support the 20 A range supported by the single HCSMU
operation.