Agilent B1500A/B1505A Programming Guide, Edition 11 4-145
Command Reference
ODSW?
Syntax ODSW chnum,state[,normal,[delay,width]]
Parameters chnum : SPGU channel number. Integer expression. 1 to 10 or 101 to 1002. See
Table 4-1.
state : 0: pulse switch disabled (initial setting)
1: pulse switch enabled
normal: 0: normally open (switch is normally open, initial setting)
1: normally closed (switch is normally closed)
delay: Only for the PG mode. Delay time (seconds) from start of pulse output
to changeover of pulse switch. Numeric expression. 0 (initial setting) to
pulse period-1E-7 seconds, setting resolution 1E-8 seconds.
width: Only for the PG mode. Duration (seconds) to hold the switched state of
the pulse switch. Numeric expression. 1E-7 (initial setting) to pulse
period-delay seconds, setting resolution 1E-8 seconds.
The delay and width parameters are ignored in ALWG mode.
Example
Statements
OUTPUT @B1500;"ODSW 101,1,1,1E-6,2E-6"
ODSW?
This query command returns the pulse switch settings for the specified SPGU
channel.
Syntax ODSW? chnum
Parameters chnum : SPGU channel number. Integer expression. 1 to 10 or 101 to 1002. See
Table Table 4-1.
Query Response state[,normal,[delay,width]]<CR/LF^EOI>
state : 0: pulse switch disabled
1: pulse switch enabled
normal : 0: normally open (switch is normally open)
1: normally closed (switch is normally closed)
delay : Delay time (seconds) from start of pulse output to changeover of pulse
switch.
width : Duration (seconds) to hold the switched state of the pulse switch