5-4 Agilent B1500A/B1505A Programming Guide, Edition 11
Error Messages
Operation Error
130 Start and stop must be same polarity.
For a log sweep, the polarity of the start and stop values must be the
same in the WV, WI, WSV, WSI, or WNX command. Also, 0 is not
allowed for the start and stop values.
140 Invalid setup
Check the setup required for the specified function and set it properly.
150 Command input buffer is full.
Agilent B1500 can receive 256 characters maximum including the
terminator at one time.
151 This command is not allowed to this channel.
152 Cannot use failed module.
The channel number specifying the module failed the self-test or
calibration. Specify another module that passed the self-test or
calibration. For the service purpose, execute the RCV command to
enable the module.
153 No module for the specified channel.
Module is not installed in the slot specified by the channel number.
154 Cannot recover this module.
The specified module cannot recover from the fail status. It may be
defective.
155 Module initialization failed.
The specified module failed the initialization. It may be defective.
160 Incorrect ST execution.
The internal memory programming can be started by the ST command
and completed by the END command. Do not enter the ST command
between the ST command and the END command.
161 Incorrect END execution.
The internal memory programming can be started by the ST command
and completed by the END command. Do not send the END command
before starting the programming.