4-230 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
WT
Enter the TM1, TM2, or TM4 command to change the mode from TM3.
Enter the OS command.
Trigger the B1500 to start measurement via the trigger input terminal.
Trigger the B1500 to recover from wait state set by the WS command via the
trigger input terminal.
Example
Statements
OUTPUT @B1500;"WSX 2"
WT
The WT command sets the hold time, delay time, and step delay time for the
staircase sweep or multi channel sweep measurement. This command is also used to
set the step source trigger delay time effective for the step output setup completion
trigger and the step measurement trigger delay time effective for the start step
measurement trigger. For the trigger function, refer to “Trigger Function” on page
2-74.
If you do not enter this command, all parameters are set to 0.
This command is not effective for the measurements using pulse.
Syntax WT hold,delay[,Sdelay[,Tdelay[,Mdelay]]]
Parameters hold : Hold time (in seconds) that is the wait time after starting the sweep
measurement and before starting the delay time for the first step.
0 to 655.35, with 10 ms resolution. Numeric expression.
delay : Delay time (in seconds) that is the wait time after starting to force a step
output and before starting a step measurement.
0 to 65.535, with 0.1 ms resolution. Numeric expression.
Sdelay : Step delay time (in seconds) that is the wait time after starting a step
measurement and before starting to force the next step output value.
0 to 1.0, with 0.1 ms resolution. Numeric expression.
If this parameter is not set, Sdelay will be 0.
If Sdelay is shorter than the measurement time, the B1500 waits until
the measurement completes, then forces the next step output.
Tdelay : Step source trigger delay time (in seconds) that is the wait time after
completing a step output setup and before sending a step output setup
completion trigger.