
2-32 Agilent B1500A/B1505A Programming Guide, Edition 11
Remote Mode Functions
Measurement Modes
If the hold time, pulse period, and trigger interval are very short, pulse settling
time may be shown.
For 0 V output, enter the DZ command that is used to memorize the present
settings of the channel and change the CMU output to 0 V for both AC and DC.
NOTE The PTDCV command sets the pulse timing parameters, such as pulse width and
pulse period.
The PDCV command sets pulsed DC bias (voltage) with AC signal.
The FC command sets the AC signal frequency.
The ACV command specifies the oscillator level, and applies the AC signal.
If the SCUU (SMU CMU Unify Unit) is connected to the CMU and two
MP/HRSMU modules correctly, the source module is automatically selected by the
DC bias setting. The CMU is selected if it is ±25 V or less (setting resolution: 0.001
V), or the SMU is selected if it is greater than ±25 V (setting resolution: 0.005 V).
Comentarios a estos manuales