Agilent B1500A/B1505A Programming Guide, Edition 11 4-215
Command Reference
WDCV
WDCV
This command sets the DC bias sweep source used for the CV (DC bias) sweep
measurement (MM18). The sweep source will be MFCMU or SMU.
Execution
Conditions
The CN/CNX command has been executed for the specified channel.
If you want to apply DC voltage over ± 25 V using the SCUU, the SCUU must be
connected correctly. The SCUU can be used with the MFCMU and two SMUs
(MPSMU or HRSMU). The SCUU cannot be used if the HPSMU is connected to
the SCUU or if the number of SMUs connected to the SCUU is only one.
If the output voltage is greater than the allowable voltage for the interlock open
condition, the interlock circuit must be shorted.
Syntax WDCV chnum,mode,start,stop,step[,comp]
Parameters chnum : MFCMU or SMU channel number. Integer expression. 1 to 10 or 101 to
1001. See Table 4-1 on page 4-16.
mode : Sweep mode. Integer expression.
1: Linear sweep (single stair, start to stop.)
2: Log sweep (single stair, start to stop.)
3: Linear sweep (double stair, start to stop to start.)
4: Log sweep (double stair, start to stop to start.)
start, stop : Start or stop value of the DC bias sweep (in V). Numeric expression.
For the log sweep, start and stop must have the same polarity.
See Table 4-7 on page 4-24, Table 4-9 on page 4-26, or Table 4-12 on
page 4-27 for each measurement resource type.
For MFCMU, 0 (initial setting) to ± 25 V (MFCMU) or ± 100 V (with
SCUU)
With the SCUU, the source module is automatically selected by the
setting value. The MFCMU is used if the start and stop values are
below ± 25 V (setting resolution: 0.001 V), or the SMU is used if they
are greater than ± 25 V (setting resolution: 0.005 V).
The SMU connected to the SCUU will operate with the 100 V limited
auto ranging and 20 mA compliance settings.
step : Number of steps for staircase sweep. Numeric expression. 1 to 1001.