Agilent B1500A/B1505A Programming Guide, Edition 11 3-15
Programming Examples
Pulsed Spot Measurements
Pulsed Spot Measurements
To perform pulsed spot measurements, use the following commands.
Function Command Parameters
Enables channels CN [chnum ... [,chnum] ... ]
Disables channels CL [chnum ... [,chnum] ... ]
Sets filter ON/OFF [FL] mode[,chnum ... [,chnum] ... ]
Sets series resistor ON/OFF [SSR] chnum,mode
Measurement time AIT 2,mode[,N]
Forces constant voltage DV, TDV chnum,range,output
[,comp[,polarity[,crange]]]
Forces constant current DI, TDI
Sets pulse timing parameters PT hold,width[,period [,tdelay]]
Sets pulse voltage PV chnum,range,base,pulse[,comp]
Sets pulse current PI chnum,range,base,pulse [,comp]
Sets voltage measurement
range
[RV] chnum,range
Sets current measurement
range
[RI] chnum,range
[RM] chnum,mode[,rate]
Sets measurement mode MM 3,chnum
Sets SMU operation mode [CMM] chnum,mode
Executes measurement XE