Agilent B1500A/B1505A Programming Guide, Edition 11 1-3
Programming Basics
Before Starting
Before Starting
Before starting the programming using the Agilent FLEX command, perform
following.
1. Terminate the Agilent EasyEXPERT software as follows.
a. Select File > Exit on the EasyEXPERT main window.
b. Click [x] at the upper right corner of the Start EasyEXPERT button.
2. Open the Agilent Connection Expert window by clicking Agilent IO Control
icon on the Windows task bar and selecting Agilent Connection Expert.
3. Change the following setup items as shown below. The setup window can be
opened by highlighting GPIB0 in the Instrument I/O on this PC area, and
clicking Change Properties... button.
GPIB address B1500’s GPIB address (ex: 17)
System Controller No
Auto-discover No
The factory shipment initial values are 17, No, and No, respectively.
4. Reboot Required dialog box is opened, reboot the B1500.
NOTE Start EasyEXPERT button
Leave the Start EasyEXPERT button on the B1500 screen. The button must be
displayed on the screen or minimized to the Windows task bar. The Start
EasyEXPERT service must be run to control the B1500 from an external computer.