Agilent Technologies B1500A Manual de servicio Pagina 168

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 598
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 167
2-82 Agilent B1500A/B1505A Programming Guide, Edition 11
Remote Mode Functions
Trigger Function
To Receive Measurement Trigger
To use an external trigger just for starting measurement, instead of the XE
command, perform the next step. This is not effective for the high speed spot
measurement.
1. Connect a BNC cable between the Ext Trig In connector and a trigger output
connector of an external device.
2. Create a control program. Then the TM3 command and HP BASIC ENTER
statement should be entered as shown in the following example:
:
OUTPUT @B1500;"MM1" ! Sets spot measurement mode
: ! Sets measurement condition
:
OUTPUT @B1500;"TM3" ! Uses external trigger
ENTER @B1500 USING "#,3X,12D,2X";M_data
:
3. Execute the control program.
The B1500 sets the measurement conditions, and waits for an external trigger
(negative trigger) sent to the Ext Trig In connector.
When the trigger is received, the B1500 starts measurement. When
measurement is completed, the B1500 sends a negative edge trigger to the Ext
Trig Out connector, and puts the measurement data in the data output buffer.
NOTE The HP BASIC ENTER statement pauses program execution until measurement
data is put in the data buffer, reads the data from the buffer, and then continues
program execution.
Start measurement
Measurement completion
Ext Trig In
Ext Trig Out
Vista de pagina 167
1 2 ... 163 164 165 166 167 168 169 170 171 172 173 ... 597 598

Comentarios a estos manuales

Sin comentarios