4-64 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
CORRDT
Parameters chnum : MFCMU channel number. Integer expression. 1 to 10 or 101 to 1001.
See Table 4-1 on page 4-16.
corr : Correction data to measure. Integer expression. 1, 2, or 3.
1: Open correction data
2: Short correction data
3: Load correction data
Query Response result<CR/LF^EOI>
0: Correction data measurement completed successfully.
1: Correction data measurement failed.
2: Correction data measurement aborted.
Example
Statements
OUTPUT @B1500;"CORR? 9,3"
ENTER @B1500;Result
CORRDT
This command sets the MFCMU open/short/load correction data to the B1500.
The correction data is cleared by the CORR? command which performs the
correction data measurement and set it to the B1500.
Syntax CORRDT chnum,freq,open_r,open_i,short_r,short_i,load_r,
load_i
Parameters chnum : MFCMU channel number. Integer expression. 1 to 10 or 101 to 1001.
See Table 4-1 on page 4-16.
freq : Frequency to be added. Numeric expression in Hz. See Table 4-18 on
page 4-30.
open_r : Open correction data (G). Numeric expression in S.
open_i : Open correction data (B). Numeric expression in S.
short_r : Short correction data (R). Numeric expression in Ω.
short_i : Short correction data (X). Numeric expression in Ω.
load_r : Load correction data (R). Numeric expression in Ω.
load_i : Load correction data (X). Numeric expression in Ω.