Agilent B1500A/B1505A Programming Guide, Edition 11 4-127
Command Reference
LSVM
Icomp: Current compliance value (in A). Numeric expression. See Table 4-7 on
page 4-24, Table 4-9 on page 4-26, Table 4-12 on page 4-27, or Table
4-15 on page 4-28 for each measurement resource type. If you do not
specify Icomp, the previous value is set. Zero amps (0 A) is not allowed
for Icomp.
Example
Statements
OUTPUT @B1500;"LSV 1,0,0,20,.5,1E-6"
LSVM
The LSVM command selects the data output mode for the linear search
measurement (MM14).
Syntax LSVM mode
Parameters mode : Data output mode. Integer expression. 0 (initial setting) or 1.
0 : Returns Data_search only.
1 : Returns Data_search and Data_sense.
Data_search is the value forced by the search output channel set by LSI or LSV.
Data_sense is the value measured by the search monitor channel set by LGI or LGV.
For data output format, refer to “Data Output Format” on page 1-25.
Example
Statements
OUTPUT @B1500;"LSVM 1"
MCC
The MCC command clears the settings of the sampling output channels defined by
the MV, MI, or MSP command. This command is available for the sampling
measurement (MM10).
Syntax MCC [chnum[,chnum...[,chnum[,chnum]]...]]
A maximum of 15 channels can be set.
Parameters chnum : Channel number of the unit to clear the settings. Integer expression. See
Table 4-1 on page 4-16.
If you do not specify chnum, the MCC command clears the settings of
all sampling channels.