4-40 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
AIT
10Value that defines the integration time given by the following
formula. N=1 to 127. Default setting is 6.
Integration time = N × initial integration time
where initial integration time is the integration time
automatically set by Agilent B1500 and you cannot change.
1 Value that defines the integration time given by the following
formula. N=1 to 127. Default setting is 3.
Integration time = N × 80 μsec
2 Value that defines the integration time given by the following
formula. N=1 to 100. Default setting is 1.
Integration time = N / power line frequency
3 Not applicable.
20Not applicable.
1 Not applicable.
2 Value that defines the number of averaging samples given by the
following formula. N=1 to 100. Default setting is 1.
Number of averaging samples = N × 128
The Agilent B1500 gets 128 samples in a power line cycle,
repeats this for the times you specify, and performs averaging to
get the measurement data.
3 Actual measurement time, in seconds. N=2 μs to 20 ms, 2 μs
resolution. Default setting is 2 μs.
For HR/MP/HPSMU, it performs the operation of type=0 and
mode=1. Then the number of averaging samples is automatically
set to the following value.
Number of averaging samples = N / Tm
Tm: Required time for one sample measurement
If N / Tm < 1, Number of averaging samples =1.