Agilent Technologies B1500A Manual de servicio Pagina 278

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 598
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 277
3-100 Agilent B1500A/B1505A Programming Guide, Edition 11
Programming Examples
C-t Sampling Measurements
C-t Sampling Measurements
To perform C-t sampling measurements, use the following commands.
Function Command Parameters
Enables Measurement Units CN [chnum ... [,chnum] ... ]
Disables Measurement Units CL [chnum ... [,chnum] ... ]
Sets Filter ON/OFF [FL] mode[,chnum ... [,chnum] ... ]
Sets series resistor ON/OFF [SSR] chnum,mode
Disables SCUU status indicator [SSL] chnum,mode
Controls SCUU input-output
path
[SSP] chnum,path
Sets MFCMU A/D converter [ACT] mode[,N]
Sets MFCMU measurement
mode
[IMP] mode
Sets timing parameters MTDCV h_bias,interval,points[,h_base]
Sets constant voltage source MDCV chnum,base,bias[,post]
Sets MFCMU output frequency FC chnum,freq
Forces AC voltage by using
MFCMU
ACV chnum,ac_level
Forces DC voltage by using
SMU
[DV, TDV] chnum,range,output
[,comp[,polarity[,crange]]]
Forces DC current by using
SMU
[DI, TDI]
Sets MFCMU measurement
range
[RC] chnum,mode
[,range]
Sets measurement mode MM 26,chnum
Executes measurement XE
Vista de pagina 277
1 2 ... 273 274 275 276 277 278 279 280 281 282 283 ... 597 598

Comentarios a estos manuales

Sin comentarios