Agilent B1500A/B1505A Programming Guide, Edition 11 2-55
Remote Mode Functions
SPGU Module
PG Operation Mode
In the PG mode (pulse generator operation mode), the SPGU outputs normal 2- or
3-level pulse voltage or DC bias voltage. To set the PG mode, execute the SIM 0
command and use the commands listed in Table 2-2 to output pulse voltage or DC
bias voltage. See Figure 2-30 for information on control commands and output
timing.
Figure 2-30 SPGU Pulse Output Control Commands and Output Timing
In the example shown in Figure 2-30, channels 801 and 802 are used for pulse
output, and channels 901 and 902 are used for DC bias output. Channels 801 and
901 output at base voltage level in the specified order when the SPUPD command is
executed. Channels 802 and 902 start their output when the SRP command is
executed in the order of their channel numbers, except for pulse outputs, which are
always started simultaneously. Also note that pulse outputs stop simultaneously on
one of the following stop conditions.
Stop conditions:
SPP command
Repeat count reaches the number specified by SPRM 1 (only for count mode)
Output time specified by SPRM 2 elapses (only for duration mode)
SPGU 901
SPGU 902
SPGU 801
SPGU 802
SMU
SPUPD 801,901 SRP
SPP
DV
base
base
base
base
base
base