4-26 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
Table 4-9 MC/HC/DHCSMU Voltage Source Setup Parameters
1
Table 4-10 HVMCU Pulsed Voltage Source Setup Parameters
2
Output
range
(actually
used)
Setting
resolution
in V
voltage, start,
stop, base,
bias, or pulse
in V
Maximum Icomp value in A
a
a. When the compliance value is set to > 1 A or < -1 A, the current measurement range is fixed to
the compliance range.
Maximum
pulse base
value
MC
SMU
HC
SMU
DHC
SMU
0.2 V 2E-7 0 to ± 0.2 ± 0.1 for DC,
± 1 for pulse
± 1 for DC,
± 20 for pulse
±2for DC,
± 40 for pulse
± 0.2 V
2 V 2E-6 0 to ± 2 ± 2 V
20 V 2E-5 0 to ± 20 ± 20 V
40 V 4E-5
0 to ± 40
b
b. Maximum ±30 V for the MCSMU.
± 1 ± 2
± 40 V
b
1. The following commands are used for setting the HCSMU, DHCSMU, and MCSMU to the
voltage source.
DV, TDV, WV, WSV, WNX, PV, PWV, LSV, LSSV, BSV, BSSV, MV, MCPNX, MCPWNX,
WDCV
Output range
(actually used)
Setting resolution
in V
start, stop, or pulse
in V
a
a. 0 A is valid for the pulse base value.
Compliance range
Maximum Icomp
value
1500 V 1.5E-3 0 to ± 1500 2 A ±2.5 A
3000 V 3E-3 0 to ± 2200 1 A ±1.1 A
100 mA ±110 mA
2. The following commands are used for setting the HVMCU to the voltage pulse source.
PV, PWV, MCPNX, MCPWNX