4-178 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
SPPER
SPPER
This command sets the pulse period for the SPGU channel. This setting applies to
all SPGU modules installed in the B1500. See “SPGU Module” on page 2-53.
Syntax SPPER period
Parameters period : Pulse period. Numeric expression. 2E-8 to 10 seconds, setting
resolution 1E-8 seconds. Initial setting 1E-6 seconds.
Example
Statements
OUTPUT @B1500;"SPPER 20E-6"
SPPER?
This query command returns the pulse period for the SPGU channel.
Syntax SPPER?
Query Response period<CR/LF^EOI>
Returns the pulse period (seconds).
Example
Statements
OUTPUT @B1500;"SPPER?"
ENTER @B1500;A
SPRM
This command specifies the operating mode for SPGU channel outputs. This setting
applies to all SPGU modules installed in the B1500.
Syntax SPRM mode[,condition]
Parameters mode : Operating mode of the SPGU channel output. Integer expression. 0, 1,
or 2. Initial setting 0.
0: Free Run mode. Continues outputting until the SPP command is
executed. The condition parameter is not required.
1: Count mode. Outputs the number of pulses (when set to PG mode
with the SIM 0 command), or the number of sequences (when set
to ALWG mode with the SIM 1 command) specified by the
condition parameter.
2: Duration mode. Outputs for a duration specified by the condition
parameter.