2-90 Agilent B1500A/B1505A Programming Guide, Edition 11
Remote Mode Functions
Initial Settings
Table 2-11 CMU Settings
Setup Item Initial Setting Commands
SCUU path/indicator open/enable SSP/SSL
Measurement parameter Cp-G IMP
Measurement range auto RC
ADC integration time auto ACT
Open/short/load correction off OPEN/SHOR/LOAD
Phase compensation mode auto ADJ
AC signal 0 V, 1 kHz ACV, FC
Sweep source parameters cleared WDCV
Automatic abort function off WMDCV
Output after measurement start value WMDCV
Hold time 0 s WTDCV
Delay time 0 s WTDCV
Step delay time 0 s WTDCV
Trigger delay time 0 s WTDCV