4-94 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
ERPFDS?
0: Disable the series resistor connection, initial setting.
1: Enable the series resistor connection.
Example
Statements
OUTPUT @B1500;"ERPFDS 1"
ERPFDS?
This query command returns the connection status for the series resistor on the
HVSMU path of Agilent N1265A ultra high current expander/fixture.
Syntax ERPFDS?
Query Response state<CR/LF^EOI>
Connection status of the series resistor on the HVSMU path. 0 (disconnect) or 1
(connect).
Example
Statements
OUTPUT @B1500;"ERPFDS?"
ENTER @B1500;A
ERPFGA
This command specifies the MC/HCSMU module connected to the Gate input of
Agilent N1265A ultra high current expander/fixture.
Execution
Conditions
No SMU may be in the high voltage state (forcing more than the allowable voltage
for the interlock open condition, or voltage compliance set to more than it).
Digital I/O port must be set to the N1265A control mode using the ERMOD 4
command.
Syntax ERPFGA gsmu
Parameters gsmu : Slot number for the MC/HCSMU connected to the Gate input.
Effective slot numbers are 1 to 10. Integer expression. Use the smaller slot number
for the HCSMU which occupies two slots.
Enter 0 to cancel the assignment.
Remarks If the HCSMU is connected to the Gate input, the maximum current of the HCSMU
for gate controlling is as follows:
DC: 100 mA