4-36 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
ACT
ACT
This command sets the number of averaging samples or the averaging time set to the
A/D converter of the MFCMU.
Syntax ACT mode[,N]
Parameters mode : Averaging mode. Integer expression. 0 (initial setting) or 2.
•0: Auto mode.
Defines the number of averaging samples given by the following
formula. Then initial averaging is the number of averaging samples
automatically set by the B1500 and you cannot change.
Number of averaging samples = N × initial averaging
• 2: Power line cycle (PLC) mode.
Defines the averaging time given by the following formula.
Averaging time = N / power line frequency
N : Coefficient used to define the number of averaging samples or the
averaging time. Integer expression.
•For mode=0: 1 to 1023. Initial setting/default setting is 2.
•For mode=2: 1 to 100. Initial setting/default setting is 1.
Example
Statements
OUTPUT @B1500;"ACT 0,1"
OUTPUT @B1500;"ACT 2,2"
ACV
This command sets the output signal level of the MFCMU, and starts the AC
voltage output. Output signal frequency is set by the FC command.
Execution
Conditions
The CN/CNX command has been executed for the specified channel.
Syntax ACV chnum,voltage
Parameters chnum : MFCMU channel number. Integer expression. 1 to 10 or 101 to 1001.
See Table 4-1 on page 4-16.
voltage : Oscillator level of the output AC voltage (in V). Numeric expression.