Agilent B1500A/B1505A Programming Guide, Edition 11 4-91
Command Reference
ERMOD?
ERMOD?
This query command returns the control mode for the digital I/O ports.
Syntax ERMOD?
Query Response mode<CR/LF^EOI>
0: General purpose control mode
1: 16440A SMU/PGU selector (B1500A-A04) control mode
2: N1258A/N1259A control mode
4: N1265A control mode
8: N1266A control mode
16: N1268A control mode
Remarks If some control modes are enabled, the ERMOD? command returns sum of mode
values. For example, after the ERMOD 4, ERMOD 8, and ERMOD 16 commands
are executed, the ERMOD? command returns 28 (=4+8+16).
Example
Statements
OUTPUT @B1500;"ERMOD?"
ENTER @B1500;A
ERPFDA
This command specifies the measurement resources connected to the Selector Input
of Agilent N1265A ultra high current expander/fixture.
Execution
Conditions
No measurement resource may be in the high voltage state (forcing more than the
allowable voltage for the interlock open condition, or voltage compliance set to
more than it).
Digital I/O port must be set to the N1265A control mode using the ERMOD 4
command.
Syntax ERPFDA hvsmu,smu
Parameters hvsmu : Slot number for the HVSMU connected to the Selector Input HVSMU
port directly or via the N1266A expander.
smu : Slot number for the MP/HPSMU connected to the Selector Input SMU
port.