4-106 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
FMT
Example
Statements
OUTPUT @B1500;"FMT 1"
OUTPUT @B1500;"FMT 2,1"
Table 4-24 FMT format parameter
12 digits data will be sn.nnnnnEsnn, snn.nnnnEsnn, or snnn.nnnEsnn.
13 digits data will be sn.nnnnnnEsnn, snn.nnnnnEsnn, or
snnn.nnnnEsnn.
where, s is + or -, E is exponent symbol, and n means one digit number.
NOTE For the 4 byte binary data output format, the time stamp function is not available.
Refer to “Data Output Format” on page 1-25.
format Data format Terminator
1
a
a. Compatible with the Agilent 4142B data output format.
ASCII (12 digits data with header) <CR/LF^EOI>
2
a
ASCII (12 digits data without header) <CR/LF^EOI>
3
a
4 byte binary <CR/LF^EOI>
4
a
4 byte binary <^EOI>
5
a
ASCII (12 digits data with header) ,
11 ASCII (13 digits data with header) <CR/LF^EOI>
12
ASCII (13 digits data without header)
b
<CR/LF^EOI>
13 8 byte binary <CR/LF^EOI>
14 8 byte binary <^EOI>
15 ASCII (13 digits data with header) ,
21
ASCII (13 digits data with header)
b
b. Compatible with the Agilent 4155/4156 FLEX mode ASCII data.
<CR/LF^EOI>
22
ASCII (13 digits data without header)
b
<CR/LF^EOI>
25
ASCII (13 digits data with header)
b
,